Philips BT1 Battery Replacement
A new Philips BT1 battery has a shelf life of five (5) years stored at proper temperatures. Although a lithium battery may be stored for up to five (5) years, it self-discharges over time. Therefore, when the Philips AED battery is eventually placed in the AED, it’s useful life will be reduced depending on how long it “sat on the shelf." Depending on usage levels, three years is the common lifespan of a Philips BT1 battery. This AED battery is non-rechargeable.
Recommended Replace Interval: Three (3) years, depending upon usage. (see Useful Life and Expected Capacity below)
When optimally maintained, a new Philips BT1 battery pack has a capacity of 1.3 Amp hours, which is equivalent to 5 hours of “ON” time. It is important to know that just turning the AED on uses up battery capacity. Also, each year, the Philips BT1 AED battery capacity decreases while the battery is in the AED due to the normal self-discharge rate. After 4 years of absolutely no patient use, approximately 35% of the useful life will remain
Contact us online or toll-free at 1-877-775-4381 to talk to one of our battery specialists about the Philips BT1 replacement battery today.
The Philips BT1 battery features the following:
- Model: NCEBT1 - FDA Approved
- Version: Non-Rechargeable
- Chemistry: LiMn02
- OEM Part# M3865 Philips - Not a replacement for the FR2 or FR2+
- Fits: BT1, Heartstream Forerunner
- Volts: 18V
- Amps: 1.3
- Weight: 1.2 LBS
- Minimum Capacity: 100 Shocks